Electro Technical  
Thermal & Humidity Laboratory  
Dimension Laboratory  
Mass Volume Density  
Viscosity Laboratory  
Pressure Laboratory  
Force Torque PSV Laboratory  
Bio-medical Laboratory  
   
Electro Technical & Relay Laboratory

ELECTRO-TECHNICAL (SOURCE & MEASURE MODE – At Lab & At Site)

Sl. No. INSTRUMENTS / PARAMETERS THAT CAN BE CALIBRATED RANGE / REMARKS
1. D C Voltage - (Source Mode)
(Measure Mode)
100 µV to 1000 V
100 µV to 1000 V
2. D C Current (Source & Measure Mode)
1 µA to 1000 A
3. Resistance (2 Wire / 4 wire)
(Source Mode)
(Measure Mode)
Md to 1 Td
Md to 1 Gd
4. A C Voltage – (Source Mode)
(Source Mode)
(Measure Mode)
(Measure Mode)
50 to 10 kHz / 1 mV to 1000
10 kHz to 20 kHz / 1 mV to 300 V
50 to 10 kHz / 1 mV to 1000 V
20 kHz / 1 mV to 300 V
5. Frequency – (Source Mode)
(Measure Mode)
0.5 Hz to 600 MHz
1 Hz to 10 MHZ
6. A C Current – (Source Mode)
(Source Mode)
(Measure Mode)
(Measure Mode)
(Measure Mode)
50 Hz to 10 kHz / 33 µA to 20 A
50 Hz / 20 A to 1000 A
50 Hz to 10 kHz / 33 µA to 1 A
50 Hz to 5 kHz / 1 A to 20 A
50 Hz to 5 kHz / 1 A to 20 A
7. Phase Angle (Lead / Lag) –
(Source & Measure Mode)
0 ° to 90 °
8. Time – (Source Mode) 1 s to 9900 s
9. A C Power (1 Phase, 50 Hz) –
(Source Mode)
Volt: 0.1 V to 600 V / Current: 0.03 A to 20 A / 0.2 pf to 1 pf
Volt: 50 V to 600 V / Current: 20 A to 1000 A / 0.2 pf to 1 pf
10. A C Power (3 Phase, 50 Hz) –
(Measure Mode)
3 x 240 V & 110 V
3 x 1 A, 5 A, 30 A
0.2 pf to 1 pf
11. Inductance – (Source Mode)
(Measure Mode)
10 µH to 10 H @ 1 kHz
100 µH to 1 H @ 100 Hz, 1 kHz
12. Capacitance – (Source Mode)
(Measure Mode)
100 pF to 90 µF @ 1 kHz
100 pF to 1000 µF @ 100 Hz, 1 kHz
13. Oscilloscope 1 mV to 100 V / 2 ns to to 5 s / <= 600 MHZ
14. Temperature Indicator / Controller / Recorder / Transmitter – RTD.
K , J, E, T, N, R, S, B – (Source & Measure)
- 200 °C to 1800 °C
15. Micro Ohm Meter / Contact Resistance Meter / Winding Resistance Meter – (Source Mode) 50 µd to 20 kd / 7 mA to 200 A
16. Power Factor (Lead/Lag) 0.2 pf to 1 pf
17. A C High Voltage – (Source Mode)
D C High Voltage
1kV to 25 kV / 50 Hz
1kV to 50 kV
18. A C Energy – 1 phase @ 50 Hz
(Measure Mode)
240 V / 1 A, 5 A, 30 A (0.5 pf to UPF)
19. A C Energy – 1 phase @ 50 Hz
(Measure Mode)
240 V / 1 A, 5 A, 30 A (0.5 pf to UPF)
20. A C Energy – 3 phase @ 50 Hz
(Measure Mode)
240 V / 1A, 5 A, 30 A (0.5 pf to UPF)
110 V / 5A (0.5 pf to UPF)

Standard Instrument used in Electro-Technical Laboratory:
Multifunction Calibrator, Make: Fluke, Model: 5520A
Multifunction Calibrator, Make: Fluke, Model: 5080A
Reference Multimeter (8 ½ Digit) , Make: Fluke, Model – 8508A
Precision Digital Multimeter (7 ½ Digit), Make: Time Electronics, Model: TE 5075
Programmable Electronic Timer, Make: DCES, Model: D1014
Frequency Calibrator, Make D.C.E.S, Model No. D1019
Phase Angle Calibrator, Make : D.C.E.S, Model No: D-1021
AC/DC Current Probe , Make: GOSSEN METRAWATT
High Voltage Divider with kV Meter, Make: Udeyraj, Model: VD-100
High Voltage Probe, Make: Fluke, Model: 80K-40
Accuchek LT+, Make: Secure Meters Ltd.
Accuchek HT+, Make: Secure Meters Ltd.
Digital LCR Meter, Make: Aplab, Model: MT4080A
Digital Megger, Make: AVO International, Model No: BMM 2000
Digital Time Calibrator, Make : DCES(P)L, Model No.: D1018
Digital Power Meter, Make: Yokogawa, Model No. 760401
Lux Meter, Make: Metravi, Model No. 1333
Sound Level Calibrator, Make: Lutron, Model No. SC-942
Digital Tachometer Make: Electronic Automation, Model DT-2001-B
High Stability Decade Megohm Box, Make: Vaiseshika, Model No. 8400
High Resistance Decade Box, Make: Tinsley, Model: 4721
High Precision Decade Resistance Box, Make: Vaiseshika, Model: 7400
Decade Inductors, Make Omega, Model No: DL-120P
Decade Inductance Box, Make: Zeal, Model No: ZMDIB
Decade Capacitance Box, Make: Time Electronics, Model No: 1071
Digital Earth Resistance Tester, Make: Kusam-Meco, Model: KM-1520
Micro ohm Calibrator (Ductor Cal), Make: Time Electronics, Model: 5070
Three Phase Power Meter Field Calibrator, Make : HPU, Model : HPU-3030B

TESTING OF PROTECTION RELAYS:
EMC is actively involved in the testing of protection relays through direct orders and also through tender process.

We test protection relays (NUMERICAL AND ELECTRO MECHANICAL) of the under mentioned categories:
Over Current
Earth Fault
Over Voltage
Under Voltage
Frequency based
Power Factor based
Fuse protection
Auxiliary
Time based
Motor Protection
Transformer Protection
Differential and many more of single phase, three phase and poly-phase variety.

Our universal Relay Test Kit from ISA (Italy) enables us to calibrate relays of all famous brands like:
ENGLISH ELECTRIC
SHNEIDER
ALSTHOM
AREVA
MICOM
MOTPRO
L&T
SIEMENS and many more.

Download E-Brochure of Electro Meter Corporation

  © Copyright © 2019-20. Electro Meter Corporation - All rights reserved.
Find us on Facebook Find us on Twitter Find us on Google+